Convert every pressure instrument specification into kilopascals at the pressures the plant actually uses. A percentage of span gives a fixed error allowance on a fixed range. A percentage of reading changes with pressure and can include a fixed term. Which instrument has the narrower allowance can change halfway across a duty cycle.
Compare the two offers at three pressures
An illustrative purchasing brief covers a 0–1,000 kPa gauge pressure transmitter. Offer A quotes ±0.15% of its 1,000 kPa span. Its error limit is ±1.5 kPa throughout the range under the stated reference conditions. Offer B quotes ±(0.4% of reading + 0.4 kPa). At 100 kPa that is ±0.8 kPa. At 500 kPa it is ±2.4 kPa. At 800 kPa it is ±3.6 kPa. Both offers are hypothetical. The fixed term in B matters particularly at low pressure, where a reading-only percentage could suggest an unrealistic zero error at zero pressure.
The crossover follows from equating the magnitudes. Solve 1.5 = 0.004 × pressure + 0.4 and the result is 275 kPa. Below 275 kPa B has the smaller quoted limit. Above it A has the smaller limit. Ranking the two offers by the percentage printed in large type would miss this change. A process normally at 100 kPa and a test bench normally at 800 kPa can make different purchasing decisions from the same two datasheets.

Separate the instrument limit from the plant decision
Suppose an Australian wash skid needs the transmitter’s contribution to stay within ±1.25 kPa near 100 kPa. From the two stated specifications, only B fits that allocation at this operating point. If the plant also needs an allocation of ±2 kPa at 800 kPa, only A fits at the higher point. Neither offer satisfies the combined brief as written. The buyer can request a better specification, split the measurement duties or revisit the allocation with the process owner. Choosing the cheapest of two individually plausible models does not resolve the two-point requirement.
Check what the specification covers. A reference accuracy at a nominated ambient temperature can exclude temperature effects, long-term drift, installation and readout contributions. A transmitter with user-adjustable output endpoints may still express its tolerance against its sensor’s upper range limit. For a bipolar instrument, full scale and full span can differ. Ask the manufacturer for the denominator and the conditions in words, especially when the quoted range crosses zero. The arithmetic above uses a unipolar range with an explicit span of 1,000 kPa.
Ask for evidence a second engineer can read
Give the pressure reference, gauge or absolute, and the points important to control or acceptance. Include tolerances in kPa for each point, plus the expected ambient and process temperatures.
Ask for readings on increasing and decreasing pressure, along with the reference instrument, its uncertainty, the test date, and the applicable specification. This supports a check for direction-dependent differences.
Keep the sign of each observed error. A 101.2 kPa indication against a 100.0 kPa reference gives +1.2 kPa. Rounding both to the nearest 5 kPa would conceal that offset.
Agree the decision rule if the certificate must state compliance. The manufacturer’s tolerance and the uncertainty of the calibration are different quantities. NATA’s equipment assurance guidance explains why the decision rule matters for covered accredited activities.
Sources and working data
Fluke’s guide to evaluating pressure specifications explains reading-based and fixed-range expressions. NATA’s equipment assurance guidance provides the Australian calibration context. Download the worked specification comparison. The supplier publishing desk includes a separate example of how to submit measured readings and a data dictionary.