Where to find accredited calibration in Australia and New Zealand

Start with NATA or IANZ, then check the scope against the instrument, location and measurement points. A coolroom logger shows which questions belong in the quote.

Industry Applied Editorial Published 2 min read

Three example temperature calibration points for a coolroom logger at zero, five and eight degrees Celsius.

A dispatch team has a coolroom logger and needs calibration at 0 °C, 5 °C and 8 °C. The first search result says the laboratory is accredited. The next question is whether its published scope covers the complete measurement the team wants. You can make most of that check before asking for a price.

Start with the national directory

In Australia, open NATA’s Find an Accredited Organisation and search for temperature calibration. In New Zealand, use IANZ’s directory and look for calibration laboratories. Open the individual facility record and its current scope or schedule. Record the laboratory, its accreditation identifier and the service location. A company with several locations may list different services at each.

Read the scope as a buyer

  • Match the quantity, instrument type and range. Calibrating an electrical simulator or a readout alone leaves a different part of the measurement chain untested from calibrating a connected logger and probe together.

  • Check that the relevant part of the scope covers each requested temperature. If the logger must stay in the coolroom, ask whether the scope and quoted method cover work at your premises.

  • Tell the laboratory the tolerance you use for this job. Ask what measurement uncertainty it expects at the requested points. If you need a pass/fail statement, agree how that decision will account for uncertainty before accepting the quote.

From a calibration need to a comparable quote. Instrument Logger and probe. Scope check 0, 5 and 8 °C. Quote request Range and uncertainty.

What to put in the enquiry

For the example logger, supply the logger and probe identifiers, their normal arrangement, the three test points and the return address. Ask for results as received, and for as-left results if an adjustment is authorised. Ask the lab to identify any work outside the accredited scope, confirm whether it will issue an endorsed report for the requested work, and quote turnaround including freight. Record any on-site access or downtime constraints.

ISO/IEC 17025 provides the international framework for laboratory competence. The national directories are where a buyer can check a particular facility and its authorised scope today. Save the scope alongside the quote so the person receiving the calibration report can reconcile the two.

Directories and guidance

NATA — Find an Accredited Organisation

NATA — industry user guide to calibration laboratories (PDF)

IANZ — directory of accredited organisations

IANZ — metrology and calibration laboratories

ISO — ISO/IEC 17025 overview