Model the source resistance in series with a 100 kilohm input and calculate the resulting voltage divider. At 10 on the source resistance (kω) scale, the worked result is 4.545 on the logged voltage from a 5 v source (v) scale. The figure and table let you check the result against a second operating point.

Set up the field measurement

Check the actual input impedance on the selected measurement range and the sensor output circuit. Repeat a known steady voltage measurement with and without the logger connected. Keep the reference observation alongside the calculation, with the instrument identification, observation period and units. This makes it possible for the next shift to distinguish a new operating condition from a change in how the observation was recorded.

Two operating points

Compare the two adjacent inputs of 5 and 10 kω. The calculation gives 4.762 at the first and 4.545 at the second. The signed change is -0.2165 on the output scale. A spreadsheet cross-check should reproduce both rows and the direction of change before it is copied through the remaining conditions.

Logged voltage from a 5 V source (V) plotted against source resistance (kω). Across the example the output changes from 4.995 to 3.333.
Logged voltage from a 5 V source (V) plotted against source resistance (kω). Across the example the output changes from 4.995 to 3.333. Figure by Industry Applied.

Worked data

Source resistance (kΩ)

Logged voltage from a 5 V source (V)

0.1

4.995005

1

4.950495

5

4.761905

10

4.545455

20

4.166667

50

3.333333

Checking the working range

At the two ends of the example, an input of 0.1 gives 4.995 and an input of 50 gives 3.333. Use a value near each end to check your implementation. Testing only the middle row can conceal an incorrect offset, an inverted ratio or a unit conversion applied twice.

Interpreting the result

This is steady-state resistive loading. Settling problems on a multiplexed input can persist even if the static voltage-divider error is small. If measured performance sits outside the worked range, first repeat the comparison using field inputs and the same boundary. Preserve the original reading in the export. A discrepancy is easier to investigate when the field observation and the calculated output occupy separate columns.

Source and calculation data

For the underlying measurement context, see NI’s data-acquisition and signal-conditioning guide. The operating values and sensitivity analysis here are an Industry Applied worked example. Download the input and output values to verify the calculation with the values from your own site.