Two irrigation meters can agree on the conductivity of a water sample and display different total dissolved solids values. First compare the conductivity readings in the same units and at the same temperature reference. Then compare the conversion factors the instruments use for their TDS display. An electrical conductivity conversion is a change of units. A conversion from conductivity to estimated mass concentration introduces an assumption about the dissolved ions.
Put both conductivity readings on one scale
A reading of 1.2 dS/m is the same conductivity as 1.2 mS/cm or 1,200 µS/cm. A decisiemens per metre and a millisiemens per centimetre each represent 0.1 siemens per metre. The numbers match in those two units. A microsiemens per centimetre is one thousandth of a millisiemens per centimetre. Confusing the m and µ prefixes can move a farm dashboard reading by three orders of magnitude without any change in the water.
Write the reference temperature alongside the units. Conductivity depends on temperature and some instruments report a compensated value referred to 25°C. A reading at sample temperature and a value adjusted to a standard reference need the settings and sample conditions checked before comparison. Keep raw water temperature with the original measurement where the meter makes it available. It becomes particularly valuable when a bore sample is taken cold and tested again after warming on a bench.

Why two ppm numbers can both be plausible
Take the same 1,200 µS/cm sample. An instrument configured to multiply µS/cm by 0.50 displays an estimated 600 mg/L on that scale. A second instrument using 0.64 displays 768 mg/L, and a 0.70 factor gives 840 mg/L. The difference between the first and third estimates is 240 mg/L although their electrical measurements are identical. NSW’s salinity guidance includes a commonly used 0.64 conversion and explains that the relationship varies with the dissolved salts.
A meter’s ppm label often refers to its estimated TDS scale. In this worked freshwater example the numerical comparison uses the familiar dilute-water approximation between mg/L and ppm by mass. For concentrated solutions, water density and the instrument’s stated reporting convention matter. A meter’s calculated number is also a different measurement from a laboratory’s gravimetric dissolved-solids result. Agree which measure a contract, agronomist or irrigation assessment actually requests.
Work through a disagreement on the farm
At a hypothetical Riverland irrigation property, a handheld meter shows 768 ppm while the pump-station dashboard shows 1.20. Before comparing the numbers, inspect the field names. If the dashboard is in dS/m and the handheld uses the 0.64 TDS scale, they represent the same example. Converting both displays to a single conductivity basis resolves the apparent conflict.
If the conductivity values still differ, draw a representative sample for both instruments, inspect and clean the cells in accordance with their instructions, and check the calibration standard and temperature settings. Record probe identity, sample point, time, temperature, units and reference settings. Keep the original observations before changing a scaling factor in the dashboard. A correction to the display must remain distinguishable from a change in bore water.
Worked data
Conductivity (µS/cm) | Conductivity (dS/m) | TDS factor | Estimated TDS (mg/L) |
|---|---|---|---|
1200 | 1.20 | 0.50 | 600 |
1200 | 1.20 | 0.64 | 768 |
1200 | 1.20 | 0.70 | 840 |
The three tabled estimates represent alternative factors for one water sample. Download the conductivity and TDS comparison. NSW explains salinity measurement units and the limits of the ppm estimate. Its water salinity testing note provides further Australian field guidance.