What does a 0.1 mm display step contribute to measurement uncertainty?

Convert a rounding interval into a standard uncertainty under a stated model. The extra decimal place is valuable only if other errors let it matter.

Industry Applied Editorial Published 2 min read

Two digital indicators examining reference blocks. AI-generated editorial illustration.
AI-generated editorial illustration by Industry Applied · Industry Applied editorial artwork

A digital gauge that rounds to the nearest 0.1 mm can display the same value for positions spread across nearly 0.1 mm. If all positions within a display bin are treated as equally plausible, the rounding contribution to standard uncertainty is the step divided by √12. For a 0.1 mm step this gives about 0.0289 mm.

Keep half-width and standard deviation distinct

The half-width of the rounding interval is 0.05 mm. Treating that as a standard deviation overstates the contribution in this rectangular model. A 0.01 mm display similarly contributes about 0.00289 mm. The table shows the scale of this one term only; none of the values is a complete gauge accuracy specification.

Each value uses display increment divided by the square root of twelve. The model assumes uniform position within a rounding bin.
Figure 1. Each value uses display increment divided by the square root of twelve. The model assumes uniform position within a rounding bin. Original figure by Industry Applied Editorial. Illustrative data.

Worked data

Display step

Half-step (mm)

Rounding contribution (mm)

0.10 mm

0.050

0.02887

0.05 mm

0.025

0.01443

0.01 mm

0.005

0.00289

Download the figures in this worked example (CSV)

Is an upgrade worth investigating?

Suppose a separate independent positioning component already contributes 0.08 mm standard uncertainty. Combining it in quadrature with 0.0289 mm gives roughly 0.0851 mm. With the finer 0.00289 mm term the combined result is about 0.0801 mm. A tenfold finer display gives a modest change in this particular measurement because fixture positioning dominates.

A device that truncates, filters aggressively or exhibits hysteresis is not described fully by nearest-step rounding. Repeated values that never change do not prove zero uncertainty. Record the display rule, repeatability and how the part is located. Use an uncertainty model appropriate to the decision and avoid counting a component twice if observed repeatability already captures it.

Sources

NIST Technical Note 1297 · Law of propagation of uncertainty

NIST/SEMATECH · Controlling bias and variability

Cover illustration by Industry Applied, generated with AI. Two digital indicators examining reference blocks.